Model | IM-6125 | |
Type | Measurement head | |
Image sensor | 1″ 6.6 mega pixel CMOS | |
Display | 10.4″ LCD monitor (XGA: 1024 ? 768) External monitor connectable (clone output) |
|
Receiver lens | Double telecentric lens | |
Image measurement | Field of view | Wide field: o100 ? L200 mm, High precision: 25 ? L125 mm |
Minimum display unit | 0.1 µm | |
Repeatability | W/o stage movement | Wide field: ±1µm High precision: ±0.5µm |
With stage movement | Wide field: ±2µm High precision: ±1.5µm | |
Measurement accuracy (±2?) | W/o binding | Wide field: ±5µm*1 High precision: ±2µm*2 |
With binding | Wide field: ± (7 +0.02L)µm*3 High precision: ± (4 +0.02L)µm*4 | |
External remote input | No-voltage input (with contacts, without contacts) | |
External output | OK/NG/FAIL/MEAS. | Relay outputRated load 24 VDC 0.5AON resistance 50 m? or less |
Interface | LAN | RJ-45 (10BASE-T / 100BASE-TX / 1000BASE-T) |
USB 2.0 series A | 6 ports (front: 2, rear: 4) | |
Record | Hard disk drive | 250 GB |
Illumination system | Transparent | Telecentric transparent illumination (Green LED) |
Ring | Four division ring illumination (White LED) | |
External | Externally connected illumination (coaxial) | |
X stage | Moving range | 100 mm (electric) |
Withstand load | 2 kg | |
Z stage | Moving range | 30 mm (electric) |
Power supply | Power voltage | 100 to 240 VAC 50 / 60 Hz |
Power consumption | 310 mA or less | |
Environmental resistance | Ambient temperature | +10 to +35 °C |
Relative humidity | 20 to 80 % RH (No condensation) | |
Weight | Approx 31 kg | |
*1 In the range of o80 mm from the centre of the stage within the operating temperature range of +23°C ±1°C at the focused focal point position *2 In the range of o20 mm from the centre of the stage within the operating temperature range of +23°C ±1°C at the focused focal point position *3 In the range of o80 ? 180 mm from the centre of the stage, within the operating temperature range of +23°C ±1°C at the focused focal point position, and with a load weighing 1 kg or less on the stage (L = amount of stage movement in mm units) *4 In the range of o20 ? 120 mm from the centre of the stage, within the operating temperature range of +23°C ±1°C at the focused focal point position, and with a load weighing 1 kg or less on the stage (L = amount of stage movement in mm units) |
Измерительная система Keyence IM-6125
Measurement Head: Wide-field model
Описание
Получить КП
Для получения коммерческого предложения, пожалуйста, вышлите ваш запрос на электронную почту [email protected] с указанием требуемых позиций и реквизитами компании.
Категория: Измерительные системы